PMMA/ Anthracene Film as a Low Doses Dosimeter

Main Article Content

Razzak Hameed

Abstract

A prepared PMMA/Anthracene film of thickness 70μm was irradiated under reduced pressure ~10-3 to 60Coγ-ray dose of (0.1mrad-10krad) range. The optical properties of the irradiated films were evaluated spectrophotometrically. The absorption spectrum showed induced absorption changes in the 200-400nm range. At 359nm, where there is a decrease in radiation-induced absorption, the optical density as a function of absorbed dose is linear from 10mrad-10Krad.It can therefore, be used as radiation dosimeter for gamma ray in the range 10mrd-10krad

Article Details

How to Cite
1.
PMMA/ Anthracene Film as a Low Doses Dosimeter. IJP [Internet]. 2009 Oct. 1 [cited 2024 Apr. 29];7(9):8-12. Available from: https://ijp.uobaghdad.edu.iq/index.php/physics/article/view/890
Section
Articles

How to Cite

1.
PMMA/ Anthracene Film as a Low Doses Dosimeter. IJP [Internet]. 2009 Oct. 1 [cited 2024 Apr. 29];7(9):8-12. Available from: https://ijp.uobaghdad.edu.iq/index.php/physics/article/view/890

Similar Articles

You may also start an advanced similarity search for this article.