Quality assurance of the linear accelerator device using Star Track and Perspex
Main Article Content
Abstract
In this study, the quality assurance of the linear accelerator available at the Baghdad Center for Radiation Therapy and Nuclear Medicine was verified using Star Track and Perspex. The study was established from August to December 2018. This study showed that there was an acceptable variation in the dose output of the linear accelerator. This variation was ±2% and it was within the permissible range according to the recommendations of the manufacturer of the accelerator (Elkta).
Article Details
This work is licensed under a Creative Commons Attribution 4.0 International License.
Authors retain copyright and grant the journal right of first publication with the work simultaneously licensed under a Creative Commons Attribution (CC-BY) 4.0 License that allows others to share the work with an acknowledgement of the work’s authorship and initial publication in this journal.